<h2>Darkfield Microscopy</b>
 +
<p>Dark-field microscopy permits the detection of unstained small biological objects which
 +
otherwise provide insufficient contrast under transmitted light observation. In a dark-field
 +
microscope a special condenser or aperture stop is used to produce light rays that normally miss
 +
being collected by the microscope objective’s collection NA. When this light interacts with a
 +
scattering object such as a biological sample, light is scattered into the collection NA of the
 +
objective and detected (Figure 2). While it is often the least used of the contrast modes you are
 +
exploring, it is relatively simple to adapt a standard microscope for darkfield imaging, and the
 +
images produced can be quite striking.</p>
 +
<p>The aperture stop in a darkfield condenser causes light to focus on the specimen in a ‘hollow’
 +
cone. The aperture stop is sized in such a way that this hollow cone is larger than the collection
 +
NA of the objective being used, so most of the light put out by the lamp is not detected. Any
 +
light that interacts with the sample will be scattered or refracted into the collection NA of the
 +
objective being used (light grey region) and detected. Light that does not pass through the
 +
sample does not enter the objective’s collection volume, and is thus not detected (gold region).</p>
 +
 +
<table width=400 align=center><td>
 +
<p align=justify>[[File:Fig2_darkfield.png|400px|border|center]]
 +
<b>Figure 2 -</b> Darkfield illumination path in a typical microscope. An aperture is used to create a ‘hollow’ beam of light
 +
and is focused at the sample. Light that does not interact with the sample (scattering, refraction, reflection, etc) is
 +
outside of the lens NA and is therefore not detected. Light that is scattered or refracted by the sample enters the
 +
collection NA of the objective and is detected.
 +
<br clear=right>
 +
</p>
 +
</td></table>
 +
<p>

Revision as of 14:18, 2 September 2010

Required Components

  1. Diatom slide
  2. Stage micrometer slide
  3. Stained and unstained baboon eye slides
  4. Fresh onion
  5. Knife and tweezers
  6. Toothpick or popsicle sticks
  7. Blank slides and coverslips
  8. Syringe of vaseline with pipet tip end
  9. Distilled water
  10. Immersion oil
  11. Nikon upright microscope and CCD camera
  12. USB stick for transferring data and images for processing and report generation

Objectives

Students will get a refresher on the optical light path of microscopes as well as working with digital acquisition software and data processing. Building off of previous knowledge, the students will explore light contrast modes for unstained samples like darkfield imaging and phase contrast imaging. In the first part of this lab these modes will be compared with standard brightfield illumination on stained and unstained specimens. In the second part darkfield imaging will be used to measure the cytoplasmic streaming rate of spherosomes in living onion cells.

Introduction

Imaging has become a vital tool for researchers in virtually all aspects of modern biophysics. Recent advances in microscope technology as well as labelling techniques and gene and protein manipulation methods have led to breakthroughs in our understanding of biological processes. In order to take advantage of these methods you, the biophysicist, need to understand some of the fundamental techniques and concepts in microscopy. You already have experience with microscopes, and this lab will build on that knowledge and get you familiar with more advanced methods as well as working with acquisition software.

The brightfield (transmitted light) imaging mode is something you have already been exposed to. Brightfield imaging mimics the human optical system and is measured in terms of colour and light intensity absorbed by a sample. The contrast method is based off of colour specific absorption of light by the dyes added into the specimen. When working with certain samples, particularly live specimens, the addition of exogenous contrast agents isn’t always feasible as they can potentially interfere with the organism or cells under observation. In this lab we will focus on some specialized microscope techniques to look at unstained samples in a way that lets you observe their morphology without the addition of foreign compounds. The imaging modes you will be using are:

  1. Brightfield: where samples contain little natural contrast and dyes are added to impart artificial colour
  2. Darkfield: where photons observed are mostly those that have been scattered by structures within the samples
  3. Phase Contrast: where optics will be used to visualize changes in the optical path length of samples

A refresher on optical microscopy and imaging theory is included in Appendix 1. Please refer to this section to review some of the fundamental concepts in microscopy. If you feel comfortable with this content you do not have to read through it thoroughly. It has been included as optional background material.

Brightfield Staining

Most biomedical microscopy specimens contain very little intrinsic contrast when viewed directly with transmitted light. For example, to make specimens such as the tissue sections you are using in this lab easier to visualize, stains and dyes are used. These compounds impart additional contrast and enable discrimination of morphological features based on their differential colour absorption across the visible spectrum.

By far, the most common type of stain used in biological microscopy is Haematoxylin & Eosin (H&E). Haematoxylin stains basophilic chemical structures such as DNA in the nucleus dark purple, while Eosin will stain the cytoplasm and extra-cellular regions a shade of pink. When oxidized, haematoxylin forms a structure called haematein. Haematein is a compound that forms strongly coloured complexes with metal ions, most notably Fe(III) and Al(III), and it is these coloured complexes which give the blue/purple colouring seen in the nuclei of H&E sections. H&E staining allows pathologists to observe the density and shape of nuclei in tissue sections, and is the gold standard in the diagnosis of many types of Cancer.

Shown in Figure 1 is a sample of an H&E stained tissue section from a rabbit brain. A low resolution image shows a brain tumour as the dark purple mass, and the zoom view shows the border between tumour and healthy tissue. Note the dark purple staining of the cancerous nuclei, and how densely packed they are.

Fig1 stained.png

Figure 1 - H&E stained section of mouse brain. Shown inset is a zoomed view showing the cluster of nuclei in a brain tumour (dark purple patch in larger image).

Darkfield Microscopy

Dark-field microscopy permits the detection of unstained small biological objects which otherwise provide insufficient contrast under transmitted light observation. In a dark-field microscope a special condenser or aperture stop is used to produce light rays that normally miss being collected by the microscope objective’s collection NA. When this light interacts with a scattering object such as a biological sample, light is scattered into the collection NA of the objective and detected (Figure 2). While it is often the least used of the contrast modes you are exploring, it is relatively simple to adapt a standard microscope for darkfield imaging, and the images produced can be quite striking.

The aperture stop in a darkfield condenser causes light to focus on the specimen in a ‘hollow’ cone. The aperture stop is sized in such a way that this hollow cone is larger than the collection NA of the objective being used, so most of the light put out by the lamp is not detected. Any light that interacts with the sample will be scattered or refracted into the collection NA of the objective being used (light grey region) and detected. Light that does not pass through the sample does not enter the objective’s collection volume, and is thus not detected (gold region).

Fig2 darkfield.png

Figure 2 - Darkfield illumination path in a typical microscope. An aperture is used to create a ‘hollow’ beam of light and is focused at the sample. Light that does not interact with the sample (scattering, refraction, reflection, etc) is outside of the lens NA and is therefore not detected. Light that is scattered or refracted by the sample enters the collection NA of the objective and is detected.